Reagent compartment
•
15 rails, each of which can hold:
•
1 x Immunoassay cartridge.
•
3 x 50 mL Biochemistry Reagent.
•
6 x 20 mL (or 5 mL) Biochemistry Reagent.
•
Storage at 12-15°C whilst operating.
•
Internal storage between 12-15°C or 8-10°C* in standby mode.
* Depending on the analyser configuration
Sample volume (µL)
•
Immunoassay: from 4 to 300 µL.
•
Biochemistry: from 4 to 50 µL.
•
Programmable in steps of 0.5 µL.
Reagent volume (µL)
•
Immunoassay: from 10 to 400 µL.
•
Biochemistry: from 10 to 400 µL.
•
Programmable in units of 1 µL.
Reaction volume (µL)
•
Immunoassay : up to 500 µl
•
Biochemistry: from 180 to 550 µL.
Pipetting system
•
Pipetting reagents and samples by probe.
•
Liquid level detection by capacitance.
•
Preheating of reagents/samples.
•
Internal and external rinsing between each pipetting of sample.
Spectrophotometer
•
Linearity: Up to 3 Abs.
•
Optic path of cuvettes: 0.8 cm.
•
Spectrophotometer with interferential filter wheel:
•
6 wavelengths available: 340, 405, 500, 540, 580 and 620 nm.
•
Light source: Halogen lamp.
•
Spectrophotometer with LEDs:
•
12 wavelengths interferential filters: 340, 405, 450, 500, 540, 550,
580, 620, 650, 700, 720 and 750 nm.
•
Light sources: multiple LEDs.
Luminometer
•
Wavelengths: from 300 to 500 nm.
•
Linearity: Up to 10 Million RLU.
Reaction carousel
•
Thermo regulated at 37°C.
•
90 positions for disposable cuvettes.
•
Automatic cuvette supply by cuvette loader.
Cuvette loader
•
Loader for cube of cuvettes.
•
Contains 960 disposable cuvettes.
•
Preheated loader.
Common liquid consumables
•
IDS-iSYS System Liquid (5 litre containers).
•
IDS-iSYS Wash (10 litre containers).
•
IDS-iSYS D-Sorb (1 litre containers).
•
Immunoassay only:
•
IDS-iSYS Triggers A & B (250 mL each).
•
AP Substrate Chemiluminescence substrate (500 mL) *
* Depending on the analyser configuration
Waste collection
•
10 liters container for liquid waste posing a biological risk.
•
Solid waste (cuvettes) disposed of in re-usable container.
Power supply
•
Voltage: 100 - 240 V.
•
Frequency: 50 - 60 Hertz.
•
Maximum power consumed: 750 VA.
Operating principle
1-3- Characteristics (continued)
IDS-iSYS User Manual - Revision M1
Operating principle 1-3
8
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...