•
With TCP/IP selected:
•
Configurable options:
The sample container is set with the type sent in the field
«
Container and Specimen ID» (see Order Block, page 9).
The user can modify this container type until the sample is
processed by the analyser (blue colour code).
The sample container sent by the LIS is ignored and is set
with the container type defined for the removable rack or for
the fixed tray.
The user can modify this container type until the sample is
processed by the analyser (blue colour code).
•
Configurable options (continued):
Active for the option «Clot Detection»
When a clot is detected in a sample, results from this
sample’s profile that have already been transferred to the
LIS will be resent (see
7- Important notes, page 67
).
Active for the option «Clot Detection»
Results that have already been transferred to the LIS are
not resent when a clot is detected in a sample.
2- ASTM Compatible Version
2-8- Line of Communication (continued)
IDS-iSYS Connection Protocol - Revision N1
2- ASTM Compatible Version
18
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...