P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology
This block represents the information relating to a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Record Type
P
Fixed value P to indicate a Patient block.
Sequence Number
1
Number of the Patient block in the message starts
at 1 and is increased for each block in the
message.
Practice Assigned Patient ID
Not used.
Laboratory Assigned Patient
ID
20105863
PID.
Patient ID #
Not used.
Patient Name
BONTEMPS^Robert
Field consisting of 5 components last name, first
name, initials, su
ffi
x and title.
Only the first 2 fields are used.
When sending a control from IDS-iSYS to LIS, the
first component in the field will be used for the
name of control.
Mother’s Maiden Name
Not used.
Birth Date
19500226
Date of birth in format YYYYMMDD.
Patient Sex
M
Reference of patient values:
•
M for Man.
•
F for Woman.
•
E for Child.
•
U for others.
In the direction LIS=>IDS-iSYS:
•
M for Man.
•
F for Woman.
•
U for Child.
•
U for others.
Patient Race
Not used.
Patient Address
Not used.
Reserved Field
Not used.
3- ASTM Compatible V2 Version
3-3- The Patient Block
IDS-iSYS Connection Protocol - Revision N1
3- ASTM Compatible V2 Version
22
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...