By using these stability data, the system manages the expiry of each individual cartridge or reagent bottle
installed in the refrigerated compartment. This expiry date is established with the two following dates:
•
lot expiry date,
•
on board expiry date, defined as the use-by date for using the cartridge or the reagent bottle.
The on board expiry date d’expiration is established by a dynamic calculation based on the following data:
•
cumulative on board time,
•
on board stability,
•
in-use stability,
•
lot expiry date.
The on board expiry date (in days and hours) is initially calculated from the first installation date of the reagent
and from the on board stability supplied with the lot data. The expiry time is set at 23:59 for all reagents.
Then this date is constantly updated taking into account the cumulative on board time. This date is finally
compared with the in-use expiry date and with the lot expiry date.
The cartridge or bottle can be used up to the date and time displayed as its on board expiry date.
The date is replaced by expiry time (23:59 ) when the day of expiry date begins.
Use
3-2- Installation of Reagents (continued)
IDS-iSYS User Manual - Revision M1
Use 3-2
45