The low level connection protocol uses the standard ASTM E 1381-95 “Standard Specification for Low-Level
protocol to transfer messages between clinical laboratory Instruments and computer systems” and the high
level connection protocol uses the standard ASTM E1394-91 “Standard Specification for transferring
information between clinical instruments and computer systems”.
Three versions are available in the software and can be selected from the IDS-iSYS interface:
The
ASTM Compatible
version is the original connection protocol version, existing in the IDS-iSYS software
versions prior to V10.00.
The
ASTM Compatible V2
version o
ff
ers additional functionalities for the transfers to the LIS:
•
transfer of the traceability associated with a result,
•
transfer of messages associated with a result in a Comment block,
•
transfer of quality control results (if the option is selected).
The
ASTM Compatible V3
version o
ff
ers, in addition to the features of ASTM Compatible V2 version:
•
transfer of the operator identifiers,
•
transfer of the traceability associated with a result including information about AP Substrate and
pretreatment washer,
•
results are resent when a clot is detected on a sample,
•
transfer ofTanner stage.
This document describes the structure and construction of ASTM messages used in each version of the
connection protocol.
1- Versions of ASTM Protocol
IDS-iSYS Connection Protocol - Revision N1
1- Versions of ASTM Protocol
3
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...