3-6- Loading of Calibrators, Controls or Samples in the Sample Compartment
Access authorised (Drawer unlocked).
Sampling under way. Access denied (Drawer locked).
Access to the sample compartment is controlled. if the light is green, access is available. When assays are
under way, the sample compartment is locked (red light).
!
When the sample compartment is opened during the run cycle, new tests will be delayed.
The reagents will continue to be aspirated for the tests which are already running.
Once the compartment is closed, the analyser can resume the sample workload,
i
ncluding
the new tests.
Close the sample compartment as soon as the loading is finished in order to maintain the
performance of the analyser.
Any position of the sample compartment may be used to install samples, calibrators and controls. Each of
these have an associated icon shown in the table below:
Calibrator
Sample
Serum, Plasma
Control
Sample
Urine
!
Sample identifiers (barcodes or manual entries) must not contain the character «%». This
character is not recognized by the system.
Avoid the use of identifiers solely containing 12 numbers, similar to the barcode structure of
calibrators and controls provided by IDS.
!
When using the 64 position sample tray, sample cups and paediatric tubes must be placed in
the tube adapters provided with the system to guarantee the quality of sampling.
Use
3-6- Loading of Calibrators, Controls or Samples in the Sample Compartment
IDS-iSYS User Manual - Revision M1
Use 3-6
60
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...