3-12- Results Storage
When a sample is completed, it must be stored for the results to be archived in the IDS-iSYS software built in
database.
If automatic storage has been programmed in the
LOCAL SYSTEM SETTINGS
menu, the profile is stored as
soon as the profile is completed.
As soon as the position occupied by this sample is released, the profile is automatically transferred to storage.
All stored results can be viewed via the DATA menu under RESULTS STORAGE.
The results can be stored individually or in multiples.
•
Click on
then click
.
•
Select the relevant profile option to be stored (all profiles, selection, profiles without message),
then click
.
The incomplete profiles can be stored when the analyser is not carrying out an assay.
This option automatically removes the analysis requests remaining in the profiles when storing the profiles.
•
The stored profiles change to a yellow colour code.
•
As soon as a position for which the profile has been stored is released, the profile will be available in results
storage.
!
It is possible to request repeat analyses for a stored profile as long as the tube is still in the
sample compartment.
!
In the
STORAGE
menu, results can be dispayed either in standard mode or in tabular mode.
The sorting options available in the work list menu can be applied to stored results.
Use
3-12-
Results Storage
IDS-iSYS User Manual - Revision M1
Use 3-12
93
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...