3-13- Quality Control Management
The results calculated for the controls are automatically stored for each reagent lot.
The results are recorded for each test and for each lot of reagent for cumulative analysis (Levey-Jennings) and,
depending on the analytical configuration, with Westgard rules for one or two levels of control.
3-13-1- Cumulative analysis
The results calculated for the controls are stored by reagent lot with the date and time of the assay. The results
are automatically recorded for cumulative examination. The results and the statistical calculations can be
displayed for a selected period of time (60 days maximum) including mean, standard deviation, CV and charts
(control results plotted around the mean value defined for the active lot).
The control data can be displayed by analyte or active lot. By selecting from the control lot, the data of the
stored lots can be displayed.
•
In the
DATA
menu, select the following pathway:
•
In the dropdown list «Analytes», select the desired analyte.
•
The software displays the controls associated with this analyte. Then select the control in the relevant
dropdown list:
Use
3-13-
Quality Control Management
IDS-iSYS User Manual - Revision M1
Use 3-13
95
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...