3-9- Adding an Emergency (STAT) Sample
Emergency (STAT) samples can be programmed for all the sample tray positions.
If the profiles are sent through by a centralised computer system, the category of ‘STAT’ is already associated
with the profile. The priority of any sample can be raised to ‘STAT’ as long as it has not already been
incorporated into the analyser workflow (brown colour code).
Otherwise, the profile must be programmed before the sample is installed on the tray.
!
Sample identifiers (barcodes or manual entries) must not contain the character «%». This
character is not recognized by the system.
Avoid the use of identifiers solely containing 12 numbers, similar to the barcode structure of
calibrators and controls provided by IDS.
•
Click on
then click on
.
•
Enter the identifier (SID).
•
Select the container: PT (primary tube), cup, ST (secondary tube).
•
Select the type of sample: Serum/Plasma, Urine, Other (see page 69).
•
Click in front of ‘STAT’.
•
Enter the full name (optional field).
•
Then select the assays to be carried out by clicking in front of the desired analyte. When an assay is
selected, a black tick is displayed.
•
When all the analyses to be carried out have been selected, click on
.
•
If the light is green, open the sample compartment.
Removable trays
•
Install the tube on a removable tray. Slide the tray into a rail in the sample compartment until the positioning
pin is fully inserted.
•
The positions and the barcoded tubes are automatically identified by the integrated barcode reader as the
tray is inserted in the rail.
•
On the interface, the tray is displayed with the free and occupied positions. Each identified position is
labelled with the corresponding icon.
•
If the barcode label is illegible, identify the products installed (see page 69).
•
Proceed in the same way for other trays to be added.
•
Close the compartment.
•
The trays defined for automatic start are automatically added to the analyser’s workload.
•
For the others, the new assays will be added to the analyser’s workload after clicking on the corresponding
button.
Fixed tray
•
Scan the barcode with the reader located on the front face. The analyser will beep when the barcode has
been correctly read.
•
Within 10 seconds, place the sample in a free position on the rack. If the sample is recognised, the analyser
will beep a second time. On the interface, the display will indicate that the position has been identified.
•
If no second beep is heard, the position is considered to be occupied but not identified. In this case, re-start
the barcode reading and installation process.
•
If the barcode label is illegible, click on
, then select ‘Vial association with samples tray’.
•
Use the keyboard to enter the barcode identifier, then select the appropriate identifier (Sample).
•
Click on
.
•
Within 10 seconds, place the sample in a free position on the tray. The position is configured.
•
Close the drawer again.
The analyser starts initiating tests again once the drawer is locked and deals with the ‘STAT’ sample
immediately.
Use
3-9-
Adding an Emergency (STAT) Sample
IDS-iSYS User Manual - Revision M1
Use 3-9
79
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...