New features in the software version V 14.04
•
The system allows reflex testing: tests can be defined with a Supervisor level of access and can be carried
out either automatically or on request.
•
Ancillary solutions IDS-iSYS TRIGGER A & B, IDS-iSYS WASH S and IDS-iSYS DSORB are now
automatically primed when installed in the system.
•
Incomplete patient profiles can now be stored together with the completed profiles. In this case, the
analysis requests that have not been performed are automatically removed.
•
The results calculated outside the measuring range are now interpreted in relation to the limits of the
measuring range.
•
The number of available tests is now indicated on Immunology cartridges common to several analytes (eg
CCS cartridge). The minimum and the maximum numbers of available tests are displayed next to the
icon.
•
The pubertal development stage (Tanner stage) can now be defined in the patient identification tab. This
data is transferred using the connection protocol «ASTM Compatible V3».
•
The qualification profile tests are no longer displayed in the calibrations/controls window.
•
New display options are available in the calibrations/controls window.
•
The criteria for automatic validation of Immunoassay calibrations are modified: a specific limit can now be
applied to the CV calculated on the RLU of each calibrator. This CV may also be excluded from the
calibration validation criteria.
•
A new criteria is now applied to Immunoassays calibrated without master curve: it allows verifying the
di
ff
erence between the target and the calculated value of each calibrator.
•
The validity of an expired calibration can be extended one day.
•
CV calculated with raw data of each control is now displayed on the first tab of calibration for
immunoassays measured by spectrophotometer.
•
An identifier entered for a calibrator or control is recorded by the system. When this ID is used again, the
lot data are automatically associated.
•
In the
CONSUMABLES
menu, the filling of the solid waste is now displayed proportional to a maximal
capacity of 400 cuvettes.
•
Post-dilutions can be performed either automatically or on request.
•
An automatic maintenance, requested by infectious disease assays, is automatically performed when the
system is placed in standby mode.
•
In the reference tables of analytical configuration, it is no longer mandatory to define an interval that does
not generate a message.
•
The connection protocol « ASTM Compatible V3» is added:
•
transfer the identifier of the operator who validates the profile.
•
transfer the Tanner stage.
•
The system can now be connected to the centralised computer via an Ethernet link.
•
Access to the Levey-Jennings menu is simplified.
•
The function for updating the analytical configuration is now integrated to the user’s interface.
•
Biochemistry assay setup is modified: modification in step configuration (compatible with previous versions
without modifying the analytical process).
Issues in software package V14.04
•
Using Westgard Rules, when a calibration has failed, a normal WE! Error message is attached to the results. When the
calibration is then restarted and been successful, normally the WE! error message should disappear. That is not the
case in this software version, the flag is still wrongly displayed on control results but not on patient profile results
•
Using Westgard Rules, in the case of a calibration with two levels of controls violating a rule (WE! Error message), only
the first control level is displayed on the Westgard graph and in the table of values.
•
When a calibration curve is displayed, only the control level 1 is shown on the curve. Other control levels must be
selected each time to be displayed.
•
An error message displays each time a Levey-Jennings graph is displayed for a control that does not have a target
value defined. This error message must be accepted prior to select the appropriate graphic representation (mean/
deviation).
•
When restoring a high volume database, the option «Stored results» for a partial restoring cannot be used. Full
restoring or the other options for partial restoring can be used.
Preface
IDS-iSYS User Manual - Revision M1
P 3
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...