5-1-1-3- Cleaning needle exterior
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Gently clean the probe with a cloth soaked in deionized water to remove traced on the needle.
•
Then use absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl) to decontaminate the needle.
•
When the cleaning procedure is finished, verify the tightness of the needle.
•
Remove the lid support tool and close the lid (see page 36).
!
Handle the probe with care. Do not twist or bend the probe during cleaning.
5-1-1-4- Decontamination of the probe
!
This maintenance activity is managed only for systems using DSORB solution.
For other systems, this maintenance activity is not required.
•
Check that IDS-iSYS D-Sorb solution is present on the analyser.
•
Click on the corresponding
button in front of this item in the maintenance list.
•
Click
. The probe will automatically aspirate the IDS-iSYS D-Sorb solution.
•
When maintenance is completed without any error, the date and time it is carried out are recorded.
Maintenance
5-1- Daily Maintenance (continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-1
117
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...