Analog output
HYDAC Filtersysteme GmbH
en
Page 72
BeWa CS1000 3247149l en.doc
2006-07-24
11.4.4 NAS 2 / NAS 5 / NAS 15 / NAS 25
The NAS x setting enables the value of a class to be continuously outputted via the analog output.
11.4.5 NAS + T
The NAS+T signal consists of 5 measured values which are transmitted time-coded with the following
time slices:
11.4.5.1 Time-coded
Signal
4,0
2,0
I (mA)
1
2
4
6
8
10
3
5
7
9
U (V)
time (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
3000
2,25
2,4
9,6
4,5
9,75
19,5
9,9
10,0
9,85
19,7
0,0
19,8
Ti
me
Signal
Size
Signal
duration per
pulse in ms
Current (I) / Voltage (U)
1
Identifier
2 µm
300
High / Low
2
Measured value
2 µm
3000
Current for measured value (see table 11.1.1)
3
Identifier
5 µm
300
High / Low / High / Low
4
Measured value
5 µm
3000
Current for measured value (see table 11.1.1)
5
Identifier
15 µm
300
High / Low / High / Low / High / Low
6
Measured value
15 µm
3000
Current for measured value (see table 11.1.1)
7
Identifier
25 µm
300
High / Low / High / Low / High / Low / High / Low
8
Measured value
25 µm
3000
Current for measured value (see table 11.1.1)
9
Identifier
Temperature
300
High / Low / High / Low / High / Low / High / Low / High / Low
10 Measured
value Temperature
3000
Current for measured value (see table 11.5.1)