Analog output
HYDAC Filtersysteme GmbH
en
Page 51
BeWa CS1000 3247149l en.doc
2006-07-24
11.1.4 SAE A / SAE B / SAE C / SAE D (SAE A/SAE B/SAE C/SAE D)
The SAE x setting enables the value of a class to be continuously outputted via the analog output.
11.1.5 SAE + T (SAE+T)
The SAE+T signal consists of 5 measured values which are transmitted time-coded with the following
time slices:
11.1.5.1 Time-coded
Signal
4,0
2,0
I (mA)
1
2
4
6
8
10
3
5
7
9
U (V)
time (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
3000
2,25
2,4
9,6
4,5
9,75
19,5
9,9
10,0
9,85
19,7
0,0
19,8
Ti
me
Signal
Size
Signal
duration per
pulse in ms
Current (I) / Voltage (U)
1
Identifier
SAE A
300
High / Low
2 Measured
value
SAE A
3000
Current / voltage for measured value (see table 11.1.1)
3
Identifier
SAE B
300
High / Low / High / Low
4 Measured
value
SAE B
3000
Current / voltage for measured value (see table 11.1.1)
5
Identifier
SAE C
300
High / Low / High / Low / High / Low
6 Measured
value
SAE C
3000
Current / voltage for measured value (see table 11.1.1)
7
Identifier
SAE D
300
High / Low / High / Low / High / Low / High / Low
8 Measured
value
SAE D
3000
Current / voltage for measured value (see table 11.1.1)
9
Identifier
Temperature
300
High / Low / High / Low / High / Low / High / Low / High / Low
10 Measured
value
Temperature 3000
Current / voltage for measured value (see table 11.5.1)