Error Messages
HYDAC Filtertechnik GmbH
en
Page 80
BeWa CS1000 3247149l en.doc
2006-10-13
12.3 Analog Output Error Signals
12.3.1 Analog signal time-coded
If the CS enters into an error status all following measured value signals are outputted in a specific
current (I) or voltage (U). Please refer to table 12.1 or 12.2 for the respective values for the current or
voltage of the output signal during an error status). The time coding is preserved.
Example: Error "Flow too low" (or "2 low") for the SAE output signal.
4,0
I (mA)
U (V)
t (ms)
4,8
Low
20,0
19,2
High
High
Low
4,5
4,3
4,1
19,7
19,5
300
3000
1
3
5
7
1
2
4
6
8
2,0
2,05
2,25
2,4
9,85
9,75
2,15
9,6
0,0
19,8
10,0
9,9
Ti
me
Signal
Size
Signal duration
per pulse
in ms
Current (I) / Voltage (U)
1
Identifier
SAE A
300
High / Low
2 Measured
value
SAE A
3000
4.4 mA / 2.2 V
3
Identifier
SAE B
300
High / Low / High / Low
4 Measured
value
SAE B
3000
4.4 mA / 2.2 V
5
Identifier
SAE C
300
High / Low / High / Low / High / Low
6 Measured
value
SAE C
3000
4.4 mA / 2.2 V
7
Identifier
SAE D
300
High / Low / High / Low / High / Low / High / Low
8 Measured
value
SAE D
3000
4.4 mA / 2.2 V