Analog output
HYDAC Filtersysteme GmbH
en
Page 66
BeWa CS1000 3247149l en.doc
2006-07-24
11.3.5 HDA.ISO – Analog Signal ISO to HDA 5500
(Digital Display Unit)
The HDA.ISO signal consists of 6 measured values (START / ISO 4 / ISO 6 / ISO 14 / ISO 21 / Status)
which are outputted sequentially. This presupposes that synchronization has been done with the
downstream control unit.
The signal output is as follows:
I (mA)
U (V)
t (s)
4
2
6
3
9
4.5
10.0
7
5
3.5
2.5
8
4
20
2 2 2 2 2 2 2 2 2 2 2
2 2 2 2 2 2 2 2 2 2 2
30
1
1
0
2
2
3
3
4
4
5
5
0
Time
Measured value
Signal duration
in s
Current (I) / Voltage (U)
Start signal 0
--
2
20 mA / 10 V
Pause
2
4 mA / 2 V
Signal 1
> 2 µm
2
Current (I) / Voltage (U) for signal (see table 11.3.5.1)
Pause
2
4 mA / 2 V
Signal 2
> 5 µm
2
Current (I) / Voltage (U) for signal (see table 11.3.5.1)
Pause
2
4 mA / 2 V
Signal 3
> 15 µm
2
Current (I) / Voltage (U) for signal (see table 11.3.5.1)
Pause
2
4 mA / 2 V
Signal 4
> 25 µm
2
Current (I) / Voltage (U) for signal (see table 11.3.5.1)
Pause
2
4 mA / 2 V
Signal 5
Status
2
Current (I) / Voltage (U) for signal (see table 11.3.5.2)
Pause
30
4 mA / 2 V
Содержание CS1000 Series
Страница 97: ...Notizen Note Note ...
Страница 98: ...Notizen Note Note ...
Страница 99: ...Notizen Note Note ...