Analog output
HYDAC Filtersysteme GmbH
en
Page 53
BeWa CS1000 3247149l en.doc
2006-07-24
11.1.6.1 HDA.SAE Signal 1-4 Table
The current or voltage range is dependent on the contamination rating according to SAE=0.0 – 14.0
(resolution: 0.1 classes).
Current I
SAE class / Error
Voltage U
I< 4.00 mA
Cable break
U< 2.00 V
I = 4.00 mA
SAE 0
U = 2.00 V
I = 4.11 mA
SAE 0.1
U = 2.06 V
I = 4.23 mA
SAE 0.2
U = 2.11 V
...
...
...
I = 5.14 mA
SAE 1
U = 2.57 V
I = 6.29 mA
SAE 2
U = 3.14 V
I = 7.43 mA
SAE 3
U = 3.71 V
I = 8.57 mA
SAE 4
U = 4.29 V
I = 9.71 mA
SAE 5
U = 4.86 V
I = 10.86 mA
SAE 6
U = 5.43 V
I = 12.00 mA
SAE 7
U = 6.00 V
I = 13.14 mA
SAE 8
U = 6.57 V
I = 14.29 mA
SAE 9
U = 7.14 V
I = 15.43 mA
SAE 10
U = 7.71 V
I = 16.57 mA
SAE 11
U = 8.29 V
I = 17.71 mA
SAE 12
U = 8.86 V
I = 18.86 mA
SAE 13
U = 9.43 V
...
...
...
I = 19.77 mA
SAE 13.8
U = 9.89 V
I = 19.89 mA
SAE 13.9
U = 9.94 V
I = 20.00 mA
SAE 14.0
U = 10.00 V
If the contamination rating according to SAE is known, the current I or voltage U can be
computed:
I = 4 mA + SAE class * (20 mA - 4 mA) / 14
U = 2 V + SAE class * (10 V - 2 V) / 14
If the current I or voltage U is known, the contamination rating according to SAE can be
computed:
SAE class = (I – 4 mA)*(14/16 mA)
SAE class = (U – 2.4 V)*(14/8 V)
11.1.6.2 HDA Status Signal 5 Table
The current or voltage of the output signal (5) is dependent on the status of the CS 1000 as shown in
the table below:
Current I
Status
Voltage U
I = 5.0 mA
The CS is functioning correctly
U = 2.5 V