Analog output
HYDAC Filtersysteme GmbH
en
Page 52
BeWa CS1000 3247149l en.doc
2006-07-24
11.1.6 HDA.SAE – Analog signal SAE to HDA 5500
(Digital Display Unit)
The HDA.SAE signal consists of 6 measured values (START / SAE A / SAE B / SAE C / SAE D /
Status) which are outputted sequentially. This presupposes that synchronization has been done with
the downstream control unit.
The signal output is as follows:
I (mA)
U (V)
t (s)
4
2
6
3
9
4.5
10.0
7
5
3.5
2.5
8
4
20
2 2 2 2 2 2 2 2 2 2 2
2 2 2 2 2 2 2 2 2 2 2
30
1
1
0
2
2
3
3
4
4
5
5
0
Time
Measured
value
Signal duration
in s
Current (I) / Voltage (U)
Start signal 0
--
2
20 mA / 10 V
Pause
2
4 mA / 2 V
Signal 1
SAE A
2
Current (I) / Voltage (U) for signal (see table 11.1.6.1
Pause
2
4 mA / 2 V
Signal 2
SAE B
2
Current (I) / Voltage (U) for signal (see table 11.1.6.2)
Pause
2
4 mA / 2 V
Signal 3
SAE C
2
Current (I) / Voltage (U) for signal (see table 11.1.6.2)
Pause
2
4 mA / 2 V
Signal 4
SAE D
2
Current (I) / Voltage (U) for signal (see table 11.1.6.2)
Pause
2
4 mA / 2 V
Signal 5
Status
2
Current (I) / Voltage (U) for signal (see table 11.1.6.2)
Pause
30
4 mA / 2 V