Installation and Operation
HYDAC Filtersysteme GmbH
en
Page 40
BeWa CS1000 3247149l en.doc
2006-07-24
60
Set test cycle time 1
– 1440 minutes
SINGLE
Perform single
measu stop
o.k.
NO SET
ANA.OUT
Select measured variable for analog output (see also section 11)
ANaOUT
Select measured
variable for analog
output
+
Designation
NASMAX
NAS maximum
NAS
NAS class 2/5/15/25 (code)
NAS+T
NAS class+temp. (code)
TEMP
Fluid temperature
HDaISO
ISO for HDA 5500
HDaNAS
NAS or SAE for HDA 5500
ISO 2
ISO 2 code
ISO 5
ISO 5 code
ISO 15
ISO 15 code
ISO
ISO 3-digit (code)
ISO+T
ISO 3-digit + temp. (code)
NAS 2
NAS class 2
NAS 5
NAS class 5
NAS 15
NAS class 15
NAS 25
NAS class 25
9.3 Description of the Switching Output in the Measuring Modes
For a more detailed description of the measurement modes, see also section 8.
9.3.1 Mode M1: Measure continuously
Purpose: Measurement
only
Function:
Continuous measurement of solid contamination, without any switching functions
9.3.2 Mode M2: Measure continuously and switch
Purpose: Continuous
measurement
and
triggering of signal lamps etc.
Function:
Continuous measurement of solid contamination, continuous monitoring of
programmable threshold; if threshold is exceeded, the switching output is activated to
monitor an alert on site
9.3.3 Mode M3: Filter down to contamination level and stop
Purpose:
Perform a single clean-up of hydraulic reservoir
Function:
Control of a filter unit, continuous measurement of solid contamination; if pre-
programmed cleanliness level is reached 5 times in sequence, the pump is stopped