2. Frequency Span Accuracy
SPECTRUM ANALYZER
SYNTHESIZED SWEEPER
ADAPTER
AD
Figure 2-4. Wide Span
Setup
10. Press
PRESET
)
on HP 8340A Synthesized Sweeper.
11. Set the synthesized sweeper to a 4
CW signal and power level
of -10
12. Press
on the analyzer.
13. Set spectrum analyzer as follows:
FREQUENCY
) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
SPAN
)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..50 0
MHZ
14. Set the synthesized sweeper to 3.8 MHz. (See
2-4.) Press
SEARCH).
15. Press MARKER
and
[
PEAK SEARCH
).
16. Press MARKER A and set the synthesized sweeper to 4.2
(See Table 2-4.) Press
SEARCH
).
17. The MARKER A frequency should be between 388 MHz and 412
MHz.
18. Set spectrum analyzer
SPAN
)
and
(
CENTER FREQUENCY
)
according to Table 2-4 and measure the frequency span by the
procedure of steps 13 through 16. The limits for the difference
between the two frequency measurements are given in the table.
Performance Tests
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...