![HP HP 8566B Скачать руководство пользователя страница 94](http://html.mh-extra.com/html/hp/hp-8566b/hp-8566b_adjustment-manual_161563094.webp)
SPECTRUM ANALYZER
16. Harmonic and Intermodulation Distortion
30. Set the other synthesized source RF OUTPUT switch to ON and
adjust the output power level for a power meter indication of
-25.00
Set both synthesized source RF OUTPUT
switches to the ON position (power meter reading should be
approximately -22
31. Connect output of power divider to analyzer RF INPUT as shown
in Figure 2-29.
Be careful to flex the cable assemblies as little as possible, as flexing
can cause a change in the measured power level. To minimize flexing,
place the power sensor close to the analyzer input.
32. Press
on the spectrum analyzer.
POWER NETER
POWER
SENSOR
ADAPTER
3
POWER
ATTENUATOR
D I V I D E R
Figure 2-29. Intermodulation Distortion
Setup
33. Key in analyzer settings as follows:
[
CENTER FREQUENCY
]
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MHZ
STEP
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MHz
FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2
34. Wait for completion of the sweep (asterisk should not appear on
display), then press MARKER
(
PEAK SEARCH
),
Wait for
completion of the sweep.
35. Press MARKER
Activate
[
CENTER FREQUENCY
]
and press
once to tune to the third order product at 2098.5 MHz.
Performance Tests 2-69
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...