4. Final Display Adjustments (SN 3004A and Above)
R 5 1 2
R 5 1 3
R 3 1 9
R 4 2 6
R 4 3 7
Figure 3-32. Location of Final Display Adjustments on
3-6. Initial Adjustment Positions
Adjustment
Position
R120 Y GAIN
centered
R127 Y POSN
centered
R220 X GAIN
centered
R227 X POSN
centered
R319 INT GAIN
two-thirds clockwise
R409 FOCUS COMP centered
R426 T/B FOC
centered
Al A2 R427
CTR
centered
R437
FOC
centered
Al A2 R440 R/L CTR
centered
R512 ORTHO
centered
R513 3D
centered
R516 INT LIM
fully counterclockwise
R517 ASTIG
centered
Front-panel INTENSITY
fully counterclockwise
Front-panel FOCUS
centered
Front-panel ALIGN
centered
4. Set the spectrum analyzer’s LINE switch to ON, and wait at least
5 minutes to allow the CRT and high-voltage circuits to warm up.
5. Set the front panel INTENSITY control fully counterclockwise and
adjust
INT LIM until the display is just visable. See
Figure 3-32.
6. Set the front-panel INTENSITY control fully clockwise.
7. Adjust
X GAIN,
X POSN,
Y
GAIN, and
Y POSN for optimum centering of the
display annotation and graticule pattern.
8. For an initial coarse focus, adjust the following potentiometers in
the sequence listed:
FOCUS LIMIT
3-66 Adjustments
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...