9. 6
Resolution Bandwidth Adjustments
24. Press MARKER
then MARKER
25. Using the DATA knob, adjust the marker down one side of the
displayed signal to
the
6
point; CRT MKR annotation indicates
x.
26. Adjust
10
for MKR A indication of 5.00
while
maintaining the marker at
X using the DATA knob. Refer to
Figure 4-7 for the adjustment location.
27. Press MARKER
Adjust the marker to the 6
point
on
the
opposite side of the signal (CRT MKR A annotation indicates 1.00
28. The CRT MKR A annotation now indicates the 6
bandwidth of
the 10
bandwidth filter. The bandwidth should be 10.0 fl.O
29. Key in
3
[
FREQUENCY
5
and
If
necessary,
readjust by pressing
(
REFERENCE
and using the DATA knob to place the signal peak near the top of
the graticule.
30. Press MARKER
and MARKER
31. Using the DATA knob, adjust the marker down one side of the
displayed signal to the 6
point; CRT MKR A annotation
indicates
X.
32. Adjust
3
for MKR A indication of 1.5
while
maintaining the marker at
X using the DATA knob. Refer to
Figure 4-7 for the adjustment location.
33. Press MARKER la]. Adjust the marker to the 6
point on the
opposite side of the signal (CRT MKR A annotation indicates 1.00
34. The CRT MKR [al annotation now indicates the 6
bandwidth
of the 3
bandwidth filter. The bandwidth should be 3.00
4-26 Option 462
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...