13. Sweep, DAC, and Main Coil Driver Adjustments
Al6
A19
SCAN
DAC
GENERATOR
TOP VIEW
A19
Al6
Figure 3-58. Location of Sweep and DAC Adjustments
10. Adjust
SWEEP TIME adjustment for sweep ramp of 500
ms duration (not including dead time at beginning and end of
each ramp) as measured on the oscilloscope.
11. Press
12. Connect the oscilloscope channel 1 to
(AUX OUT). Display
should be a stepped series of 4 sweep ramps similar to that shown
in Figure 3-59.
16 Adjustments
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...