18. RF’ Module Phase Lock Adjustments
25. Adjust
BALANCE for a dead time between
consecutive voltage ramps of 3.5 0.1 msec. Each voltage ramp
should have a “flattened” top and reach a maximum voltage
of approximately
Vdc. The minimum voltage between
consecutive voltage ramps should be approximately -24.4 Vdc.
26. Set the spectrum ananlyzer LINE switch to STANDBY. Disconnect
the oscilloscope from the RF Section
VCO TUNE.
Disconnect cable 0 (black) from
CAL OUTPUT, the cable
85 (gray/green) from
100 MHz INPUT, and cable 5 (green)
from
3.3
INPUT.
27. Remove the
Phase Lock Assembly from the extender board,
and remove the extender board from the RF Section. Reinstall the
Phase Lock Assembly in the RF Section.
28. Disconnect cable 84 (gray/yellow) from
and the cable 82
(gray/red) from
Replace the cover to the A6 RF Module,
and then reconnect cable 84 (gray/yellow) to
and cable
82 (gray/red) to
29. Reconnect cable 2 (red) to
300 MHz OUTPUT, cable 0
(black) to
CAL OUTPUT, cable 85 (gray/green) to
100 MHz INPUT, cable 4 (yellow) to
VCO TUNE, and cable
5 (green) to
3.3
INPUT on the
Phase Lock
Assembly.
30. Perform adjustment procedure 19, “CAL Output Adjustment”.
Adjustments 3-165
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...