7. 3 MHz Bandwidth Filter Adjustments
3 MHz Bandwidth Filter
Figure 3-38.
Location of Center, Symmetry, and 10 Amplitude
Adjustments
9. Remove crystal filter bypass network near Cl4 SYM.
10. Adjust
CTR for minimum amplitude of the displayed
signal peak. Adjust
SYM for best symmetry of the
displayed signal. Repeat adjustments to ensure that the signal
is
and adjusted for best symmetry. See Figure 3-38 for
location of adjustments.
11. Remove crystal filter bypass network near C23 SYM.
12. Adjust
CTR for minimum amplitude of the displayed
signal peak. Adjust
SYM for best symmetry of the
displayed signal. Repeat adjustments to ensure that signal is
and adjusted for best symmetry. See Figure 3-38 for
location of adjustments.
13. Remove crystal filter bypass network near C32 SYM.
14. Adjust
CTR for minimum amplitude of the displayed
signal peak. Adjust
SYM for best symmetry of the
displayed signal. Repeat adjustments to ensure that signal is
and adjusted for best symmetry. See Figure 3-38 for
location of adjustments.
15. Remove crystal filter bypass network near C41 SYM.
16. Adjust
CTR for minimum amplitude of
the
displayed
signal peak. Adjust
SYM for best symmetry of the
displayed signal. Repeat adjustments to ensure that the signal
is
and adjusted for best symmetry. See Figure 3-38 for
location of adjustments.
17. Signal should be centered on center graticule line on CRT display.
If signal is not centered, repeat steps 3 through 16 to readjust
each filter stage.
3-78 Adjustments
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...