The procedures in this section are for the adjustment of the
instrument’s electrical performance characteristics.
The procedures require access to the interior of the instrument
and therefore should only be performed by qualified service
personnel. Refer to
Considerations
in this introduction.
1. Low-Voltage Power Supply Adjustments . . . . . . . . . . . . . . . . . . . . .
2. High-Voltage Adjustment (SN 3001A and Below) . . . . . . . . . . . . . 3-31
2. High-Voltage Adjustment (SN 3004A and Above) . . . . . . . . . . . .
3. Preliminary Display Adjustments (SN 3001A and Below) ....
3. Preliminary Display Adjustments (SN 3004A and Above) .... 3-56
4. Final Display Adjustments (SN 3001A and Below) . . . . . . . . . . . .3-64
4. Final Display Adjustments (SN 3004A and Above) . . . . . . . . . . . .3-66
5. Log Amplifier Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-70
6. Video Processor Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-75
7. 3 MHz Bandwidth Filter Adjustments . . . . . . . . . . . . . . . . . . . . . . . . .3-78
8. 21.4 MHz Bandwidth Filter Adjustments . . . . . . . . . . . . . . . . . . . . .
9. 3
Bandwidth Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-92
10. Step Gain and 18.4 MHz Local Oscillator Adjustments ...... 3-96
11. Down/Up Converter Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . 3-102
12. 10 MHz Standard Adjustment (SN 2637A and Below) .....
12. 10 MHz Standard Adjustment (SN 2728A and Above) ..... .3-110
13. Sweep, DAC, and Main Coil Driver Adjustments . . . . . . . . . . . 3-114
14. 100 MHz VCXO Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15. M/N Loop Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-133
16. YTO Loop Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-137
17.
Loop Phase Lock Adjustments . . . . . . . . . . . . . . . . . . . . . . . 3-148
18. RF Module Phase Lock Adjustments . . . . . . . . . . . . . . . . . . . . . . . .3-161
19. CAL Output Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-167
20. Last Converter Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-171
21. Frequency Response Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . .3-176
22. Analog-To-Digital Converter Adjustments . . . . . . . . . . . . . . . . . . 3-207
23. Track and Hold Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-210
24. Digital Storage Display Adjustments . . . . . . . . . . . . . . . . . . . . . . .
The adjustment procedures should
not
be performed as routine
maintenance, but only when Performance Tests cannot meet
specifications. Before attempting any adjustment, allow the
instrument to warm up for one hour.
3-l is a cross reference of
Function Adjusted to the related Adjustment procedure. Table 3-2
lists all adjustable components by name, reference designator, and
function.
Adjustments 3-1
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...