2 1. Frequency Response Adjustments
LOG SCALE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 dB/DIV
187. Press MARKER
and adjust (IF Step Gain) potentiometer
GF for a marker indication of 0.00
fO.O1
If
Miscellaneous Bias/Relay Driver Assembly is HP
85660-60322 (HP
RF Sections with
number prefix
2747A or above), and
GF does not have sufficient range,
change the
of factory-select component
Increase
the
of
to decrease the signal level, and decrease the
value of
to increase the signal level. See
3-3 for the
acceptable range of
for
and
3-4 for HP part
numbers. See Figure 3-90 for the location of
188.
189.
190.
191.
192.
Disconnect the low-loss microwave test cable from the spectrum
analyzer front panel 321.4 MHz IF INPUT connector. Reconnect
the jumper cable between the spectrum analyzer front panel
32 1.4 MHz IF INPUT and IF OUTPUT connectors.
On the spectrum analyzer, key in
6 Hz,
(
REFERENCE LEVEL
)
-12
In the RF Section, disconnect cable 84 (gray/yellow) from
and cable 82 (gray/red) from
Replace
the cover to the A6 RF Module, and then reconnect cable 84
(gray/yellow) to
and cable 82 (gray/red) to
Remove the jumper between
and
(LOCK
INDICATOR DISABLE) on the Al2 Front Panel Interface
Assembly in the RF Section.
Replace the RF Section bottom cover.
Adjustments
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...