3. Preliminary Display Adjustments (SN 3004A and Above)
Figure 3-24. Location of
and
T P 5 0 1
R 1 2 7
C l 0 9 T P 1 0 5 R 2 2 7
GND
Figure 3-25.
Adjustment Locations
10. Set the pulse/funtion generator controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..pulse
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Width (WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250 ns
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
V
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .O.OOO
Disable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off (OUTPUT enabled)
11. Connect
the
output of the pulse/function generator to (X input)
on the Display Adjustment PC board in the
slot as shown in
Figure 3-23.
The pulse/function generator output must be terminated with 50
ohms. Use a BNC tee, a 500 termination, and a BNC female to SMB
female adapter. Install the 500 termination as close to the Display
Adjustment PC Board as possible.
3-58 Adjustments
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...