Last Converter Adjustments
Displayed average noise level (0
input attenuation, 10 Hz
resolution bandwidth):
Non-Preselected
100 Hz to 50
50
to 1 MHz
1 MHz to 2.5
Preselected
2.0
to 5.8
5.8
to 12.5
12.5
to 18.6
18.6
to 22
The displayed average noise level is measured in a 10 Hz bandwidth at
various frequencies with no signal applied to the analyzer input.
50 Ohm Coaxial Termination . . . . . . . . . . . . . . . HP
Option 012
1. Press
2. Connect CAL OUTPUT to RF INPUT.
3. Press
Adjust AMPTD CAL for a MARKER amplitude
of -10.00
f0.02
4. Disconnect CAL OUTPUT. Terminate RF INPUT with a 50 ohm
coaxial termination.
5. Press
Key in settings as follows:
START
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..8 0 Hz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180 Hz
REFERENCE LEVEL
]
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -70
VIDEO BW
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hz
6 .
Press (SHIFT)
Wait about 2 minutes for the VID AVG
readout to reach 10 sweeps or more, then press [SHIFT), TRACE B
7. Press MARKER
Tune to 100 Hz, or the nearest
frequency that is not on the slope of the LO feedthrough or on a
line-related sideband (e.g., 120 Hz). Refer to Figure 2-26.
2-58 Performance Tests
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...