2 1. Frequency Response Adjustments
22. On the synthesized sweeper, press
LEVEL
]
and adjust
the
ENTRY knob as necessary for a power meter indication of
-15.00
at 100 MHz.
23. On the power meter, press
(
RANGE HOLD
)
(turning it on).
24. On the synthesized sweeper, press
(
POWER LEVEL
)
and adjust the
ENTRY knob for a power meter indication of -10.00
f0.03
at 100 MHz.
25. On the synthesized sweeper, press
LEVELING and adjust
the ENTRY knob (REF in
with ATN: 0
for a power
meter indication of -10.00
at 100 MHz.
Do not vary the synthesized sweeper POWER LEVEL setting
(internal leveling) or METER REF and METER ATN settings (external
power meter leveling) for the remaining steps in this section of the
adjustment procedure. The frequency response adjustments are
referenced to the -10.00
power level at 100 MHz.
26. Press
on the spectrum analyzer. Set the RF Section
front-panel AMPTD CAL control to
the
approximate center of its
adjustment range.
Do not vary
the
spectrum analyzer front-panel AMPTD CAL control
setting for the remaining steps in this adjustment procedure.
27. Set (Band A Lower Segment)
Al and (Band A Upper
Segment)
A2 each to the approximate center of its
adjustment range.
Then, adjust
GAIN fully clockwise for maximum
RF gain. See Figure 3-90 for the locations of
Al,
A2, and
GAIN.
28. On the spectrum analyzer, key in
8,
[
REFERENCE LEVEL
)
-4
and then press the
and
keys as necessary to
position the peak of the displayed 100 MHz signal within one
division of the top graticule line. Key in MARKER
(
PEAK SEARCH
),
MARKER
on the spectrum analyzer to position two markers
on
the peak of the displayed 100 MHz signal.
29. Readjust
GAIN counterclockwise to decrease the RF
gain 5.00
as indicated by the MARKER A indication
on the spectrum analyzer display.
30. On
the
spectrum analyzer, key in (RECALL) 8.
31. Adjust
potentiometer (Band A Step Gain)
GA to adjust the amplitude of the displayed 100 MHz signal to
10.00
10
Adjust
GA counterclockwise
to increase the signal level, and clockwise to decrease the signal
level. If
GA does not have sufficient range, adjust the
amplitude of the 100 MHz displayed signal as close as possible to
-10.00
Adjustments
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...