None
for sweep times
seconds
The triangular wave output of a function generator is used to
modulate a 500 MHz signal which is applied to the spectrum analyzer
RF INPUT. The signal is demodulated in the zero span mode to display
the
triangular waveform. Sweep time accuracy for sweep times
ms is tested by checking the spacing of the signal peaks on the
displayed waveform.
SPECTRUM
SYNTHESIZED SWEEPER
Figure 2-35. Fast Sweep Time Accuracy
ms)
Setup
Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3312A
Universal Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A
Signal Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8340A
1. Connect equipment as shown in Figure 2-35.
2. Press
on spectrum analyzer.
3. Key in analyzer settings as follows:
(
CENTER FREQUENCY
]
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500
MHZ
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..10 0
4. Set synthesized sweeper for an output frequency of 500 MHz and
an output power level of -10
2-84 Performance Tests
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...