20. On the spectrum analyzer, key in the following settings:
(
CENTER FR
E
QUENCY
) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
100 MHz
FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21. On the spectrum analyzer, press
(
REFERENCE LEVEL
]
and use the
DATA knob to position the signal peak near the reference level
(top graticule line).
22. On the spectrum analyzer, key in
(
PEAK SEARCH
],
MARKER
and press
several times to position the second marker at the
leftmost graticule line.
23. Adjust
300
for a MARKER A indication of -3.00
f0.05
24. On the spectrum analyzer, press
several times to position
the second marker at the rightmost graticule line. Then, press
and use the DATA knob to adjust the centering of the
displayed signal for a MARKER A indication of -3.00
f0.05
25.
26.
On the spectrum analyzer, key in
(
FREQUENCY
300
MARKER
(OFF), (
PEAK SEARCH
],
MARKER
and press
several
times to position the second marker at the leftmost graticule line.
Readjust
300
for a MARKER A indication of -3.00
f0.05
27. Repeat steps 24 through 26 as necessary until no further
adjustment is required.
9. 3
Bandwidth Adjustments
The 100
bandwidth is set with factory-select components
and
If it is necessary to
increase the 100
bandwidth, increase the value of one or
more of these factory-select components. The 30
bandwidth
is set with factory-select components
and
If it is necessary to increase the 30
bandwidth, decrease the value of one or more of these
select components. Refer to
3-3 for the acceptable range of
values for
and
and to Table 3-4 for HP
part numbers.
28. On the spectrum analyzer, key in the following settings:
(
CENTER FREQUENCY
)
.....................................
100
MHZ
FREQUENCY SPAN] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
29. On the spectrum analyzer, press
[
REFERENCE LEVEL
)
and use the
DATA knob to position the signal peak near the reference level
(top graticule line).
30. On the spectrum analyzer, key in
(
PEAK SEARCH
],
MARKER (a,
and press
several times to position the second marker at the
leftmost graticule line.
Adjustments 3-91
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...