9. Impulse Bandwidth Adjustments
13. Using the DATA knob, adjust the marker down one side of
the display signal to the 7.3
point; CRT MKR A annotation
indicates 0.430 X.
I F C O N T R O L
\
Figure 4-8. Location of Bandwidth Adjustments
14. Adjust
1 MHz for MKR A indication of 500
while
maintaining the marker at 0.430 X using the DATA knob. Refer to
Figure 4-8 for the adjustment location.
15. Press MARKER
Adjust marker to the opposite side of the
signal (CRT MKR A annotation indicate 1.00 X). There are now
two markers; one on each of the signal at the 7.3
point.
16. The CRT MKR A annotation now indicates
the
impulse bandwidth
of the 1 MHz bandwidth. The impulse bandwidth should be 1.00
MHz
MHz.
17. Key in
300
500
and
If necessary, readjust by pressing
[REFERENCE LEVEL]] and using the DATA knob to place the
signal peak at the top of the graticule.
18. Press MARKER
then MARKER
19. Using the DATA knob, adjust the marker down one the displayed
signal to the 7.3
point; CRT MKR A annotation indicates 0.430
X.
20. Adjust
300
for MKR A indication of 150
while
maintaining marker at 0.430 X using the data knob. Refer to
Figure 4-8 for location of adjustment.
21. Press MARKER
Adjust the marker to the 7.3
point on the
opposite side of the signal (CRT MKR A annotation indicates 1.00
22. The CRT MKR A annotation now indicates the impulse bandwidth
of the 300
bandwidth. The impulse bandwidth should be
300.00
23. Key in
10
20
and
If necessary, readjust by pressing
4-28 Option 462
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...