X Center Frequency
+ 10
The spectrum analyzer is set to zero frequency span and the SWEEP
+ TUNE OUT auxiliary output is measured with a voltmeter as the
analyzer is tuned across its frequency range.
SPECTRUM ANALYZER
DIGITAL VOLTMETER
Figure 2-34. Sweep + Tune Out Accuracy
Setup
Digital Voltmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A
1. Press
on the analyzer. Set
(
FREQUENCY SPAN
)
to 0 Hz
2. Connect digital voltmeter to the SWEEP + TUNE OUT auxiliary
output on the rear panel of the analyzer as indicated in
Figure 2-34.
3.
Set
(
CENTER FREQUENCY
)
according to
2-22 and record the
voltmeter readings in the table. The allowable range for each
measurement is shown in the table.
2-92 Performance Tests