2-1. Performance
Cross-Reference
Function or Characteristic Tested
Test
Performance Test
No.
Center Frequency Readout
1
Center Frequency Readout Accuracy Test
Frequency Spans
2
Frequency Span Accuracy Test
Bandwidths*
3
Resolution Bandwidth Accuracy Test
Bandwidth Shape*
4
Resolution Bandwidth Selectivity Test
Bandwidth Amplitudes*
5
Resolution Bandwidth Switching Uncertainty Test
Log Scales
6
Log Scale Switching Uncertainty Test
IF Gains
7
IF Gain Uncertainty Test
Log and Linear Amplifier
8
Scale Fidelity Test
CAL OUTPUT Level
9
Calibrator Amplitude Accuracy Test
Frequency Response
10 Frequency Response Test
Sweep Times
11
Sweep Time Accuracy Test
Noise Sidebands
12 Noise Sidebands Test
Line-Related Sidebands
13 Line-Related Sidebands Test
Noise Floor
14
Average Noise Level Test
Residual Responses
15 Residual Responses Test
Harmonic and Intermodulation Distortion
16
Harmonic and Intermodulation Distortion Test
Image, Multiple, and Out-of-Band Responses
17
Image, Multiple, and Out-of-Band Responses Test
Gain Compression
18 Gain Compression Test
LO OUTPUT Amplitude
19
LO OUTPUT Amplitude Test
SWEEP+ TUNE OUT
20
SWEEP + TUNE OUT Amplitude Test
Fast Sweep Times
21
Fast Sweep Time Accuracy Test
ms)
Frequency Reference
22
Frequency Reference Error Test
‘For Option 462 instruments, refer to Chapter 4.
Option 857 instruments, refer to Chapter 5.
This instrument requires periodic verification of performance. The
instrument should have a complete verification of specifications at
least every six months.
2-2 Performance Tests
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...