![HP HP 8566B Скачать руководство пользователя страница 309](http://html.mh-extra.com/html/hp/hp-8566b/hp-8566b_adjustment-manual_161563309.webp)
20. Last Converter Adjustments
RF-Section:
Last Converter
A 321.4 MHz signal from a synthesized sweeper is applied to the 321.4
MHz IF input of the
Last Converter Assembly, and the 321.4
MHz
Filter is adjusted. Then, a 310.7 MHz signal is applied
to the 321.4 MHz IF input, and the 10.7 MHz Notch Filter is adjusted
to null 10.7 MHz subharmonic spurious responses.
Synthesized Sweeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP
Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8566B
15x2 Extender Board (service
. . . . . . . . . . . . .
Probe (SMB Male Bulkhead Connector) . . . . . . . . . . . . . . . . . .
Cables:
BNC to SMB Cable (2
required)
. . . . . 85680-60093
1. Set the spectrum analyzer LINE switch to STANDBY. Turn the
spectrum analyzer over to position the RF Section on top, as
shown in Figure 3-87, and remove the RF Section bottom cover.
2. In the RF Section, disconnect cable 84 (gray/yellow) from
and cable 82 (gray/red) from
Disconnect
cable 2 (red) from
300 MHz OUTPUT, cable 0 (black) from
CAL OUTPUT, cable 85 (gray/green) from
100
MHz INPUT, cable 4 (yellow) from
VCO TUNE, and cable
5 (green) from
3.3
INPUT.
3. Remove the cover from the A6 RF Module. Remove the
Phase Lock Assembly, the
Miscellaneous Bias/Relay Driver
Assembly, the
Slope Generator Assembly, and the
YTX Driver Assembly from the RF Section.
Adjustments
69
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...