2. High-Voltage Adjustment (SN 3001A and Below)
stabilize and the CRT to normalize. This
turn-on will extend
CRT life expectancy, particularly if a new CRT has just been
installed.
17. Readjust
HV ADJ for a DVM indication equal to the
voltage determined in step 15.
18. If a new CRT has just been installed do the following:
a. Set the front-panel INTENSITY control so the CRT trace is
barely visible.
b. Wait an additional 30 minutes for the CRT to normalize.
c. Readjust
HV ADJ for a DVM indication equal to the
voltage determined in step 15.
19. Set the spectrum analyzer LINE switch to STANDBY. Remove the
ac line cord from each instrument section.
20. Wait at least one minute for the MAINS power-on indicator
(red LED) to go out completely before proceeding.
21. Disconnect
the
high-voltage probe from
22. Remove the
Intensity Control Assembly from
the
IF-Display
Section and install in its place the Display Adjustment Board,
HP part number 85662-60088. Set the switch on the Display
Adjustment Board in the “down” position. (This applies
approximately
V
dc
to the front-panel INTENSITY control.)
23. Connect a calibrated
divider probe to the oscilloscope
Channel 1 input.
24. On the oscilloscope, press
to perform a soft reset.
On the oscilloscope, press
probe , select
channel 1, and use the front-panel knob to select a 10: 1 probe.
26. Set the oscilloscope controls as follows:
Press
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Press
time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Press
EDGE TRIGGER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . auto, edge
source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..l
level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
V, rising edge
Press (DISPLAY):
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on
27. On the oscilloscope press
28. Connect the oscilloscope channel 1 probe to
using a long
probe extension. See Figure 3-7 for the location of
29. Reconnect the ac line cords to each instrument section. Adjust
the front-panel INTENSITY control fully counter-clockwise, and
Adjustments 3-35
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...