17.
Loop Phase Lock Adjustments
3. Set the frequency synthesizer for a 20
0
output.
Connect the frequency synthesizer
50-ohm
OUTPUT to
using a BNC to Alligator clip adapter. Connect the ground lead of
the
adapter to the metal cover of the
VCO Assembly.
4. Connect the high-frequency active probe to the RF INPUT of
the second spectrum analyzer and to the probe power supply, as
shown in Figure 3-77.
5. Press
on the second spectrum analyzer. Connect the
active probe tip to the CAL OUTPUT signal, and press (RECALL].
6. Adjust the AMPTD CAL control on the second spectrum analyzer
for a -10.00
displayed signal, and then press
7. Set the controls of the second spectrum analyzer as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0
LOG SCALE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dB/DIV
VIDEO BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hz
8. Connect the active probe tip to the lead of A10AlL7 closest to
as indicated in Figure 3-78. The 20
signal (from
the 40
LPF) should be visible on the display of the second
spectrum analyzer.
9. Press MARKER
[
PEAK SEARCH
],
REF
on the second
spectrum analyzer to position the peak of the 20
signal at the
top graticule line. Press MARKER la] on
the
second spectrum
analyzer.
10. Set the frequency synthesizer for an output frequency of 50
Adjustments
and
are
sealed
at the factory and
normally do not require readjustment unless a component failure has
occurred. To verify their proper adjustment, skip to step 12.
11. Adjust
and
50
NULL adjustments to
minimize the amplitude of the 50
signal displayed on the
second spectrum analyzer. Refer to Figure 3-78 for the location of
and
50
NULL adjustments.
3-148 Adjustments
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...