8. Amplitude Fidelity
10. Key in analyzer settings as follows:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hz
FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MHz
11. Set the frequency synthesizer for an output power level of + 10
12. Press SCALE LIN pushbutton. Press MARKER
(
PEAK SEARCH
),
to center the signal on the display.
13. Set
SPAN
]
to 0 Hz and
to 1 Hz. Press (SHIFT),
(resolution bandwidth), MARKER [a.
14. Decrease frequency synthesizer output amplitude by 10
steps,
noting the MARKER A amplitude and recording it in column 2 of
Table 2-13.
2-13. Linear Amplitude Fidelity
0
-10.87 -9.21
- 1 0
-23.10 -17.72
2-30 Performance Tests
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...