Frequency Response Adjustments
(Includes input attenuator flatness in the 10
setting and mixing
mode gain variations, and assumes PRESELECTOR PEAK in current
instrument state.)
Flatness (20 to 30°C)
Tuned Frequency
100 Hz to 2.5
non-preselected band
2 to 12.5
preselected bands
12.5 to 18.6
preselected band
18.6 to 20
preselected band
20 to 22
preselected band
Cumulative
100 Hz to 20
100 Hz to 22
Absolute Amplitude Calibration
f1.7
f2.2
f2.2
k2.2
f0.6
Frequency response is checked across the full range of the spectrum
analyzer. In the non-preselected range from 100 Hz to 2.5
three
signal sources are used to make swept measurements: a function
generator (100 Hz to 100
a frequency synthesizer (100
to 60 MHz), and a synthesized sweeper (60 MHz to 2.5
In the
preselected bands from 2
to 22
a synthesized sweeper is
used to check the frequency response. From 100 Hz to 60 MHz, the
source flatness permits a direct display of analyzer response. Above
60 MHz, the externally
source is first characterized with
a power meter. The power sensor
Factor % switch is used to
compensate for the frequency response of the power meter.
2-32 Performance Tests
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...