![HP HP 8566B Скачать руководство пользователя страница 319](http://html.mh-extra.com/html/hp/hp-8566b/hp-8566b_adjustment-manual_161563319.webp)
2 1. Frequency Response Adjustments
Vdc Reference Supplies adjustment
-9
V affects
tracking, YTX diode bias, and Slope Generator Upper/Lower
Segment frequency breakpoints in preselected frequency bands
(Bands B, C, D, and E, 2
22
16. Adjust 17-turn potentiometer
-9 V for a DVM
indication of -9.000 fO.OO1 Vdc. See Figure 3-90 for the
location of
17. Press
on the spectrum analyzer. Set the controls of the
spectrum analyzer as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Hz
SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SINGLE
18. Use a BNC cable and a BNC (f) to
banana plug adapter to
connect the DVM to the RF Section rear-panel SWEEP+TUNE
OUT connector.
19. Adjust 25-turn potentiometer
ZERO for a DVM
indication of 0.000 fO.OO1 Vdc. See Figure 3-90 for the location
of
20. Connect the equipment as shown in Figure 3-91, with one
resistive output of the power splitter connected to the power
meter/power sensor using an APC 3.5
(m)
to Type N (f) adapter,
and the second resistive output connected to the spectrum
analyzer RF INPUT using an APC 3.5 (m) to Type N (m) adapter.
Connect the power meter rear panel RECORDER OUTPUT to the
synthesized sweeper front panel LEVELING EXT INPUT.
21. Press
PRESET
)
on the synthesized sweeper. Set the controls
of the synthesized sweeper as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 MHz
POWER LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . -9.0
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .INT
Figure 3-91.
Frequency Response Adjustments Setup (10
to 2.5
Adjustments 3-179
Содержание HP 8566B
Страница 71: ......
Страница 137: ...Test 19 1st LO Output Amplitude Steps Min Measured Max 4 1st LO OUTPUT Level 5 dBm 2 112 Performance Tests ...
Страница 169: ......
Страница 344: ......
Страница 351: ...23 Track and Hold Adjustments 25 Repeat steps 4 through 24 until no further adjustments are required Adjustments 3 211 ...
Страница 404: ...AIVI Wi AlAl ti2 W8 Figure 6 6 IF Section Front View 6 8 Major Assembly and Component locations ...