2-46
(2) Input lot number
Lot number (example: G01D-1) is indicated on the label of ink or makeup.
By inputting this lot number on Unit information screen, year and month of manufacture of
Ink or makeup is calculated within IJP, and makes it possible to manage the expiration
date. Input the lot number without hyphen. At input of lot number, fault is output under
conditions as below. (Refer to (4) Fault in terms of contents)
No.
Fault output condition at lot number input
1
Lot number is different
2
Year and month of manufacture of ink or makeup is future
date
3
The expiration date has already passed at the time of lot
number input
(3) Fault output starting date (Refer to (4) Fault in terms of contents)
(a) When lot number was input in user maintenance mode
->Output fault by using recommended expiration date
(
Ex.
)
1069K ; Year and month of manufacture : 2014/07
・
Expiration date of 1069K
:
24 months
・
Estimated replacement time after opening
:
6 months
Therefore, fault output starting date is
30 months from 2014/07
=
The last day of 2017/01
→
[2017/02/01]
(b) When lot number was input in service personnel maintenance mode
->Input date of lot Estimated time to replace is set as fault output starting
date. Expiration year and date of ink and makeup is ignored.
(
Ex.
)
JP-K69 ; Year and month of manufacture : 2014/11
Lot number input date: 2014 /12/15
・
Expiration date of JP-K69
:
24months
・
Year and month of manufacture
:
2014/11
・
Estimated replacement time after opening
:
6months
Therefore, fault output starting date is
6 months from 2014/12/15 is 2015/06/15
→
[2015/06/16]
Year and month
of manufacture of
Ink
2014/07
Start fault output
2017/02/01
Fault is NOT output
for 30 months
+α
Fault continues to output until lot
number of new inkbottle is input
These will be ignored
Lot number
input date
2014/12
Start fault output
2015/06/16
Fault is NOT output
for 6 months
Fault continues to output until lot
number of new inkbottle is input
Содержание IJ UX
Страница 1: ...Service Manual HITACHI Printer Model UX Revision July 1 2015 Version ...
Страница 2: ... Revision of UX service manual Revision Chapter Revised Page Frist Edition ...
Страница 15: ...1 9 2 External views Rear side 1 2 2 Main body internal Appearance with inner cover T removed ...
Страница 16: ...1 10 1 2 3 Print head UX E UX D B Air exhaust port ...
Страница 107: ...3 27 Circuit diagram of EZJ127A ...
Страница 108: ...3 28 Circuit diagram of EZJ127B ...
Страница 122: ...3 42 Circuit diagram of EZJ129 ...
Страница 131: ...3 51 3 3 20 Electrical Connection Diagram ...
Страница 132: ...3 52 ...
Страница 188: ...4 55 5 Reset the time of the R air filter to 0 on the Parts usage time management screen ...
Страница 301: ...7 Attached Drawing 7 1 Circulation System Diagram Circulation System Diagram UX B D 7 1 ...
Страница 302: ...Circulation System Diagram UX E 7 2 ...
Страница 303: ...7 3 Location of Solenoid valve Location of ICU joint Reducing valve Viscometer Pressure sensor Makeup filter Motor Pump ...
Страница 304: ...7 2 Electrical Connection Diagram 7 2 1 Electrical Connection Diagram UX D E 7 4 Electrical Connection Diagram UX D E ...
Страница 305: ...7 5 7 2 2 Electrical Connection Diagram UX B Basic model Electrical Connection Diagram UX B ...
Страница 308: ... Nozzle diameter 65μm 7 8 7 4 Dimensions around charge electrode and deflection electrode ...