4-30
4.10 Ink Reservoir/Makeup Reservoir Replacement Procedure
4.10.1
Ink Reservoir replacement
1
Drain ink from the ink reservoir.
Connect the recovery tube to waste bottle.
Execute “Reservoir Ink-Drainage” and drain the ink from the ink reservoir.
[Caution]
In case that the ink doesn't come out from the ink reservoir even when
"Reservoir InkDrainage" is executed, which may be caused by malfunction of the part
such as the solenoid valve or the pump, remove the ink reservoir in the step [5] below and
drain the ink into a beaker out of the ink reservoir's opening for attaching the sensor block.
At the ink drainage, put wiping paper under the beaker for possible ink dripping.
If there is still ink remaining in the ink cartridge bottle, do NOT lift the lever handle.
2
Turn off the power supply.
3
Perform “4.1.2 Access to the parts on the front side of the equipment” and draw out the
reservoir unit.
[Caution] Be careful not to catch the tubes.
4
Remove the sensor block from the ink reservoir.
5
Remove the fixing screws (2 positions) and remove the ink reservoir.
6
Lift the lever handle and remove the ink cartridge.
Fixing screws
Ink reservoir
Lever handle
Ink reservoir
Ink cartridge
Sensor block
Содержание IJ UX
Страница 1: ...Service Manual HITACHI Printer Model UX Revision July 1 2015 Version ...
Страница 2: ... Revision of UX service manual Revision Chapter Revised Page Frist Edition ...
Страница 15: ...1 9 2 External views Rear side 1 2 2 Main body internal Appearance with inner cover T removed ...
Страница 16: ...1 10 1 2 3 Print head UX E UX D B Air exhaust port ...
Страница 107: ...3 27 Circuit diagram of EZJ127A ...
Страница 108: ...3 28 Circuit diagram of EZJ127B ...
Страница 122: ...3 42 Circuit diagram of EZJ129 ...
Страница 131: ...3 51 3 3 20 Electrical Connection Diagram ...
Страница 132: ...3 52 ...
Страница 188: ...4 55 5 Reset the time of the R air filter to 0 on the Parts usage time management screen ...
Страница 301: ...7 Attached Drawing 7 1 Circulation System Diagram Circulation System Diagram UX B D 7 1 ...
Страница 302: ...Circulation System Diagram UX E 7 2 ...
Страница 303: ...7 3 Location of Solenoid valve Location of ICU joint Reducing valve Viscometer Pressure sensor Makeup filter Motor Pump ...
Страница 304: ...7 2 Electrical Connection Diagram 7 2 1 Electrical Connection Diagram UX D E 7 4 Electrical Connection Diagram UX D E ...
Страница 305: ...7 5 7 2 2 Electrical Connection Diagram UX B Basic model Electrical Connection Diagram UX B ...
Страница 308: ... Nozzle diameter 65μm 7 8 7 4 Dimensions around charge electrode and deflection electrode ...