4-19
6 Pull out connectors(6 positions) toward the front, which were disengaged in item 5
above. Then, remove the tubes (4 positions) and remove the screws (2 positions) that fix
the pump unit on the door.
(Remove the tube from the ICU joint, which is shown by dashed line. Remove the other
tubes (3 positions) by loosening the tube locks.
7 Remove two screws which are located topside. These screws are fixing ICU Fixing plate
to Pump unit.
Fixing screw
Fixing screw
ICU fixing plate
ICU fixing plate
Pump unit
Pull out connectors(6 positions)
Tubes
Tube
Fixing screws
Fixing screws
Door
Pump unit
Содержание IJ UX
Страница 1: ...Service Manual HITACHI Printer Model UX Revision July 1 2015 Version ...
Страница 2: ... Revision of UX service manual Revision Chapter Revised Page Frist Edition ...
Страница 15: ...1 9 2 External views Rear side 1 2 2 Main body internal Appearance with inner cover T removed ...
Страница 16: ...1 10 1 2 3 Print head UX E UX D B Air exhaust port ...
Страница 107: ...3 27 Circuit diagram of EZJ127A ...
Страница 108: ...3 28 Circuit diagram of EZJ127B ...
Страница 122: ...3 42 Circuit diagram of EZJ129 ...
Страница 131: ...3 51 3 3 20 Electrical Connection Diagram ...
Страница 132: ...3 52 ...
Страница 188: ...4 55 5 Reset the time of the R air filter to 0 on the Parts usage time management screen ...
Страница 301: ...7 Attached Drawing 7 1 Circulation System Diagram Circulation System Diagram UX B D 7 1 ...
Страница 302: ...Circulation System Diagram UX E 7 2 ...
Страница 303: ...7 3 Location of Solenoid valve Location of ICU joint Reducing valve Viscometer Pressure sensor Makeup filter Motor Pump ...
Страница 304: ...7 2 Electrical Connection Diagram 7 2 1 Electrical Connection Diagram UX D E 7 4 Electrical Connection Diagram UX D E ...
Страница 305: ...7 5 7 2 2 Electrical Connection Diagram UX B Basic model Electrical Connection Diagram UX B ...
Страница 308: ... Nozzle diameter 65μm 7 8 7 4 Dimensions around charge electrode and deflection electrode ...