5.7 Patient Leakage Current
Measurement Mode
────────────────────────────────────────────────────
5.7.1 Patient Leakage Current
Measurement Mode
Condition of the equipment to be measured
Power source polarity
Single fault condition
110% voltage applied
(normal polarity)
Normal polarity
Reverse polarity
110% voltage applied
(reverse polarity)
Normal polarity
Reverse polarity
NOTE
Class
equipment
Type BF applied part
Type CF applied part
Single fault condition
Single fault condition
110%
voltage
applied
110%
voltage
applied
Patient leakage current
−
−
−
−
−
−
−
5mA
−
−
−
0.05mA
Class
equipment
Type BF applied part
Type CF applied part
Single fault condition
Single fault condition
110%
voltage
applied
110%
voltage
applied
Patient leakage current
−
−
−
−
−
−
−
5mA
−
−
−
0.05mA
Internally powered
equipment
Type BF applied part
Type CF applied part
Single fault condition
Single fault condition
110%
voltage
applied
110%
voltage
applied
Patient leakage current
−
−
−
−
−
−
−
5mA
Notes 1
and 2
−
−
−
0.05mA
Notes 1
and 2
5.7 Patient Leakage Current
Measurement Mode
Patient leakage current
measurement for medical electrical equipment is
performed only for equipment with a type BF or type CF applied part.
Plug the power cord of the equipment to be measured into the auxiliary socket
for the 3155, and make a measurement by touching 9170 TEST LEADS to the
test point (applied part), and applying voltage from the optional 9196 APPLY
UNIT. For internally powered equipment, to achieve 250 V output, use the
9196 APPLY UNIT in combination with a regulated power supply. (Other
voltage application units can also be used to make measurements.)
The following combinations are measured under the patient leakage current
measurement.
・
In the patient leakage current
measurement mode, the 50
μ
A range is
not used.
・
For internally powered equipment, use wire for connecting the signal input
and output parts which are not protectively earthed to the ground.
Measurement types and their allowable values
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......