138
────────────────────────────────────────────────────
6.3 Enclosure Leakage Current Measurement Mode (for IEC/TR 60990)
────────────────────────────────────────────────────
15
Opens the measurement current set-up window. Set the measurement current.
When set to AC peak, the filter cannot be set to OFF. If set to OFF, it
automatically changes to ON1.
When set to DC, the filter cannot be set to ON2. If set to ON2, it automatically
changes to OFF.
: AC,
: DC,
: AC+DC,
: AC peak
For details, refer to Section 6.6.4, "Measurement Current Setting (for IEC/TR
60990)."
16
Opens the contact condition set-up window. Set the contact condition.
: Between the enclosure and the earth
: Between the enclosure and the power line
17
Zero adjustment function
Press this key to perform zero adjustment before measuring.
When zero adjustment is effective,
is displayed on the screen.
For details, refer to Section 6.3.3, "Zero Adjustment (for IEC/TR 60990 -
Enclosure Leakage Current)."
18
Opens the print-out window.
Prints out the maximum value displayed on the screen.
For details, refer to Section 6.6.5, "Printing out the Maximum Values."
19
Opens the save data window.
For details, refer to Section 6.6.6, "Saving the Maximum Values."
20
Exits the enclosure leakage current measurement, and returns to the initial
screen.
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......