![Hioki 3155-01 Скачать руководство пользователя страница 287](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164287.webp)
267
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Queries the maximum value.
Syntax
:MEASure:MAXimum?
Function
Returns the maximum value, its judgment, power source polarity, and
condition of the equipment to be measured, as numerical values in NR3 and
NR1 format.
The data are ordered as follows: <maximum value>, <judgment>, <power
source polarity> and <condition of the equipment to be measured>.
<Judgment>
(If the allowable value judgment is disabled, 0)
Allowable value or lower (PASS): 0
Higher than allowable value (FAIL): 1
<Power source polarity>
(If the equipment to be measured is set to the internally powered
equipment, or if measuring the low resistance, 0)
Normal polarity: 0
Reverse polarity: 1
<Condition of the equipment to be measured>
(If measuring the low resistance, 0)
Normal condition: 0
Single fault condition (open power lead): 1
Single fault condition (open ground): 2
Single fault condition (110% voltage applied): 3
Note
In order to learn the maximum value, execute this command after verifying
that the automatic measurement is finished (:AMC? command). Executing
this command during automatic measurement will obtain only the most
recent value.
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:MEASure:MAXimum?
Response
Headers ON
:MEASURE:M2.345E‑03,1,1,2
Headers OFF
+2.345E‑03,1,1,2
The power source polarity is reverse polarity, the condition of the
equipment to be measured is single fault condition (open ground), and the
maximum value is 2.345 mA. And the result of allowable value judgment is
"higher than allowable value (FAIL)."
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, an error occurs.
:MEASure:MAXimum?
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......