![Hioki 3155-01 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164036.webp)
16
────────────────────────────────────────────────────
3.1 Measurement Networks
────────────────────────────────────────────────────
3.1.3 9499 NETWORK D (Universal)
V
1. k
Ω
±
0.1%
V
0.15
μ
F
±
2%
1.5 k
Ω
±
0.1%
V
2 k
Ω
±
0.1%
Network of 2 k
Ω
(2 k
Ω
)
Network of 1 k
Ω
(1 k
Ω
)
Network of 1.5 k
Ω
and 0.15
μ
F (1.5 k
Ω
)
Standard
Safety of household and similar electrical appliances - Part 1:
General requirements (IEC 60335-1 (1991-06) + am1 (1994-11))
Applicable UL standards: (Examples: UL471, UL1310, UL1437, UL1492, etc.)
Measuring circuit
configuration/accuracy
(a) Network of 1 k
Ω
Impedance: 1 k
Ω±
0.5% (DC to 1 MHz)
(b) Network of 1.5 k
Ω
and 0.15
μ
F
Impedance: Theoretical value
±
0.5% (DC
≦
f
≦
400 Hz)
Theoretical value
±
(2%
+1
Ω
) (400 Hz
<
f
≦
1 MHz)
Cutoff frequency (at -3 dB): 707 Hz
±
15 Hz
(c) Network of 2 k
Ω
Impedance: 2 k
Ω±
0.5% (DC to 1 MHz)
Filter configuration and
characteristics
(1M
Ω
load with input protection fuse shorted)
OFF setting: High pass filter
Cutoff frequency (at -3 dB): 1805 Hz
±
27 Hz
Passband attenuation: 0 dB
±
2% (at 100 kHz)
Low-frequency attenuation: -12 dB
±
0.3 dB (at 100 Hz)
ON1 setting: (perception) 10 k
Ω
+22 nF
Passband attenuation: -12 dB
±
0.3 dB (at 100 Hz)
High-frequency attenuation: -42.8 dB
±
1 dB (at 100 kHz)
ON2 setting: (let-go) 10 k
Ω
+(20 k
Ω
+6.2 nF)//9.1 nF
Passband attenuation: -12 dB
±
0.3 dB (at 100 Hz)
High-frequency attenuation: -35.1 dB
±
1 dB (at 100 kHz)
Approved tolerances
for element
±
0.1% for resistance,
±
2% for capacitor
Fuse
250 V T0.032 AL 20 mm
×
5 mm dia.
Dimensions and mass
140W
×
35 H
×
29D mm(5.51"W
×
1.38"H
×
1.14"D), approx. 80 g(2.8 oz.)
Accessories
Instruction Manual, spare fuse
This is a universal measurement network.
Measurement network D conforms to the following standards:
Safety of household and similar electrical appliances - Part 1: General
requirements (IEC 60335-1 (1991-06) + am1 (1994-11))
Applicable UL standards:
By installing on the 3155 and making the appropriate settings from the
screen, a configuration with a network of 1 k
Ω
for 1 k
Ω
, a network of 1.5 k
Ω
and 0.15
μ
F for 1.5 k
Ω
, and a network of 2 k
Ω
for 2 k
Ω
is established.
In addition, there are many applicable standards.
Specifications
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......