![Hioki 3155-01 Скачать руководство пользователя страница 277](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164277.webp)
257
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Queries the measurement network filter.
Syntax
:CONFigure:FILTer?
Function
Returns the measurement network filter setting as <data>.
(ME equipment)
9497 NETWORK B
<data>
ON The measurement network filter has been set to the network with
frequency characteristics.
OFF The measurement network filter has been set to the network with
only 1 k
Ω
of uninduced resistance.
(Ordinary equipment)
9498 NETWORK C
ON1 The measurement network filter has been set to the perception and
reaction network.
ON2 The measurement network filter has been set to the let-go network.
OFF The measurement network filter has been set to the body impedance
network.
9499 NETWORK D
OFF1 The measurement network filter has been set to the network of
1 k
Ω
.
ON
The measurement network filter has been set to the network of
1.5 k
Ω
and 0.15
μ
F.
OFF2 The measurement network filter has been set to the network of
2 k
Ω
.
Note
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:CONFigure:FILTer?
Response
Headers ON
:CONFIGURE:FILTER OFF
Headers OFF
OFF
The measurement network filter has been set to the network with only
1 k
Ω
of uninduced resistance. (ME equipment)
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, an error occurs.
:CONFigure:FILTer?
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......