27
────────────────────────────────────────────────────
4.2 Power Cord Connection
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
WARNING
Versions of the 3155 that operate on 120 V, 200 V and 240 V of the
supply voltage are available. The 3155-01 operates on 110 to 120 V, 200
V and 240 V with the voltage selector.
Before turning on the power, make sure that the voltage of the power
supply being used matches the supply voltage indicated on the rear
panel of the unit. The maximum rated power of the 3155 is 30 VA.
The protective ground terminal of the unit is connected to a ground line
via a three-core power cord that is supplied with the unit.
In order to prevent electric shock, always connect the unit to a properly
grounded power outlet using the power cord provided.
CAUTION
・
To prevent an error, always remove the key and keep it after setting the
power supply voltage. (3155-01 only)
・
When changing the supply voltage, always turn off the power and disconnect
the power cord before beginning. (3155-01 only)
・
This unit is provided with a grounded three-core power cord that is suitable
for use with power supply conditions in the country or region in which the unit
is sold. When using this unit in any other country or region, be sure to
obtain a grounded three-core power cord which is suitable for use with the
power supply conditions in that area.
NOTE
(3155)
4.2 Power Cord Connection
The 3155 is a leakage current measuring instrument, and thererfore is only
able to operate to its full capability after it has been properly grounded.
Always use the power cord (grounded three-core power cord) that is supplied
with the unit to ground the unit.
(1) Be sure that the main power switch is turned
to OFF.
(2) Set the power voltage with the voltage selector.
(3155-01 only)
Insert the supplied voltage selector key into the
keyhole of the voltage selector, and select the
power voltage by setting it to the required
power voltage. (110 to 120/200/240 V)
Voltage selector key (2 supplied)
(3) Being sure that the power source is of the
correct voltage, connect the grounded three-core
power cord provided to the power inlet on the
back of the unit.
(4) Insert the plug into the outlet. The cord is
automatically grounded.
Содержание 3155-01
Страница 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Страница 2: ......
Страница 12: ......
Страница 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Страница 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Страница 129: ...109 5 11 System Screen ME Equipment Example of a data unit printout Example of a details printout...
Страница 135: ...115 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 6 Ordinary Electrical Equipment Measurement Network C Installed...
Страница 189: ...169 6 7 System Screen for IEC TR 60990 Example of a data unit printout Example of a details printout...
Страница 195: ...175 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 7 Ordinary Electrical Equipment Measurement Network D Installed...
Страница 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Страница 314: ...294...
Страница 322: ...302 9 3 Accuracy Tables...
Страница 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Страница 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Страница 362: ...INDEX 6 Index...
Страница 363: ......
Страница 364: ......
Страница 365: ......